Recall Z-0978-2019 ยท BTL Ventilator, Emergency, Powered (Resuscitator)

O-Two Medical Technologies, Inc.

e700 Automatic Transport Ventilator, Model Number 01EVE700

DateApr 4, 2019 (initiated Feb 12, 2019)
StatusTerminated
ReasonThere is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper screw insertion the screw sat higher than normal allowing it to contact the underside of the main board causing the screw to rub against the bottom side of the board, wearing away the insulation layer and creating a short circuit.
Root causeProcess control
ActionUrgent Medical Device Correction notification letters dated 2/7/19 were sent to customers. An additional letter was sent to customers on 3/4/19.
Quantity632
DistributionThe products were distributed to the following US states: AZ, CA, CO, FL, OH, PA, TX, and VA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K141595O-TWO E700,O-TWO E600,O-TWO E500