Recall Z-1005-2019 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Zimmer Trabecular Metal Technology, Inc.
NexGenTrabecular Metal Tibial Half Block Augment, Right Lateral/Left Medial, Size 4 5MM Thickness W/Screws Item Number: 00544800426 Product Usage: The Trabecular Metal Tibial Cone Augments are intended for use where severe degeneration, trauma, or other pathology of the knee joint indicates total knee arthroplasty. When used with the NexGen Complete Knee Solution Rotating Hinge Knee (RHK) System
| Date | Feb 1, 2019 |
| Status | Terminated |
| Reason | Potentially comingled, resulting in the product in the box potentially not matching the product on the label |
| Root cause | Mixed-up of materials/components |
| Action | Zimmer Biomet notified sent an Urgent Medical Device Recall on February 1, 2019 via email and Fed'X letter to Distributors, hospitals and surgeons, The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediately locate and quarantine affected product in your inventory, complete Attachment 1 Inventory Return Certification Form and send to CorporateQuality.PostMarket@zimmerbiomet.com. For questions contact Customer Service at 574-371-3071. |
| Quantity | 24 units |
| Distribution | Worldwide Distribution - US Nationwide Distribution in the states of AL, CA, CO, ID, MS,NE, NJ, NY, OH, OR, PA, TX, VA, WI and Foreign CANADA, JAPAN, GERMANY , NETHERLANDS |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K024161 | NEXGEN COMPLETE KNEE SOLUTION - TRABECULAR METAL AUGMENTS Implex Corp., Jan 15, 2003 |