ClearanceBrief
Recall Z-1102-2019 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Abbott Laboratories, Inc

Abbott ARCHITECT c4000 Processing Module, List No. 02P24 - Product Usage: The Abbott ARCHITECT System is intended for In Vitro diagnostic use only. The Abbott ARCHITECT Clinical Chemistry Systems are designed to perform automated chemistry tests, utilizing photometry and potentiometric technology. The ARCHITECT c4000, c8000, c16000 sample pipettor assemblies include a fluid sense/pressure monitori

DateJul 17, 2018
StatusTerminated
ReasonThere is a potential to generate incorrect results on the instrument if particular error codes are not resolved before transitioning to the Running status.
Root causeSoftware design
ActionRecall letters were issued in the U.S. 7/17/2018 via FedEx overnight mail notifying the consignee of the issue, patient impact, and necessary actions to take. Globally, primary methods of issuance were via mail, telephone, and email.
Quantity3,675 instruments
DistributionWorldwide distribution - US Nationwide and countries Afghanistan, Albania, Algeria, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bangladesh, Belarus, Belgium, Bermuda, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Burkina FASO, Canada, Cayman Islands, Chile, China, Colomb
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K980367AEROSET SYSTEM
Abbott Laboratories, Apr 1, 1998