Recall Z-1077-2019 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Beckman Coulter Inc.

The UniCel DxI 600 and 800 Access Immunoassay Systems, that are connected to an automation line. UniCel DxI 600 Catalog Numbers: A30260, A71460, A71461, A92060. Unicel DxI 800 Catalog Numbers: 973100, A71456, A71457, A84545, A25288, A25285.

DateMar 30, 2019 (initiated Dec 5, 2018)
StatusOpen, Classified
ReasonWhen racks are front-loaded onto the sample presentation unit while samples from the automation line are being processed, a scheduling error within the software can cause mishandling of samples during the creation of an aliquot and lead to incorrect results being generated. Affected samples typically yield non-reproducible falsely decreased results by up to 100%, however the potential for falsely elevated results cannot be ruled out.
Root causeUnder Investigation by firm
ActionOn 12/05/18, the firm mailed Urgent Medical Device Recall notices to all consignees. In order to prevent this error from occurring, do not load racks onto the SPU while samples are being processed from the automation line. Either of the following options can be followed to avoid the error: 1) Pause the automation line before loading of any racks onto the SPU. a) To pause the automation line and load racks on the SPU, use the following steps: i) Set the Lab Automation System (LAS) Mode to "Offline" (1) Return to the Main Menu (F9) (2) Select Configure (F8) (3) Select LIS/LAS (F6) (4) Select L
Quantity2842
DistributionUS: NJ, OH, NY, GA, FL, TN, WI, MI, MS, CA, NC, TX, PA, DC, IN, KY, IL, WA, MN, LA, ND, ID, WV, AK, AL, MA. OUS: Argentina, Australia, Bahrain, Canada, China, Croatia, Czech Republic, Egypt, Germany, Greece, India, Italy, Malaysia, Morocco, Netherlands, Nigeria, Poland, Portugal, Qatar, Saudi Arabi
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K023764UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM, MODEL 973100
Beckman Coulter, Inc., Jan 28, 2003