Recall Z-0989-2019 ยท FSY Light, Surgical, Ceiling Mounted
GETINGE US SALES LLC
Maquet XS Flat Screen Monitor Holder, PWD70+DF HD R XS32 K3 SPE, Configuration Part Number ARD268400150A Product Usage: This product is designed to support one or two medical grade flat screen(s) alone or in association with surgical lights.
| Date | Mar 7, 2019 (initiated Jan 8, 2019) |
| Status | Terminated |
| Reason | The affected flat screen monitor holders might not be compliant with grounding requirements. As a consequence, there is a risk of mild to severe/fatal electric shock under certain conditions while using the flat screen monitor holder with voltage of 100-220V. |
| Root cause | Other |
| Action | Urgent Medical Device Field Correction notification letters dated 1/8/19 were sent to customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to: -Complete and sign the enclosed Customer Response Form (Pages 4 and 5) and return to Getinge using the fax number or email address provided on the form. -A Getinge Service person will contact the person listed on the Customer Response Form (page 5) to schedule service for an update of your device, free of charge. You can also arrange a visit by a Getinge-trained or authorized service techn |
| Quantity | 6 |
| Distribution | The products were distributed to the following US states: AZ and TX. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.