ClearanceBrief
Recall Z-1118-2019 ยท DQY Catheter, Percutaneous

Arrow International Inc

Arrow Two-Lumen Central Venous Catheterization Kit with Blue FlexTip Catheter Product Code: AH-11802 (OUS) GTN: 00801902026132

DateJan 10, 2019
StatusTerminated
ReasonThe product lidstock contains a labelling error. The lidstock states the incorrect priming volume. The priming volumes printed on the extension lines are correct
Root causeError in labeling
ActionTeleflex issued Urgent Medical Device Notification to US consignee notification letter on 10th January 2019 via FedEx 2-day mail. Actions to be taken to remove product from the facilities: A: The affected customer is instructed to take the following actions: 1. Place a copy of this notification with each unit of affected product currently in your inventory. 2. After step (1) is complete, fill out the enclosed Acknowledgement Form
QuantityNone (US)
DistributionFL, NC
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K862056[CENTRAL VENOUS][MULTI-LUMEN] CATHETER W/FLEX TIP
Arrow Intl., Inc., Aug 25, 1986