Recall Z-0973-2019 ยท HRX Arthroscope
Stryker Sustainability Solutions
Stryker Sustainability Solutions (SSS) reprocessed Arthroscopic Shavers Arthroscopic shavers can be used to abrade, cut and excise tissue and bone; remove loose fragments; and, shave away debris in arthroscopic surgeries, as well as surgeries of the jaw and sinuses. Intended for use in orthopedic surgical procedures of the joints, jaw or sinuses where the cutting and removal of soft and hard tis
| Date | Jul 27, 2018 |
| Status | Terminated |
| Reason | Affected lots of reprocessed arthroscopic shavers may display the incorrect catalog number and speed settings when connected to the generator, require manual mode for use on CORE generators, or display an error message on Crossfire 1 and 2 generators which does not allow the device to be used. |
| Root cause | Software Design Change |
| Action | On July 26, 2018, Stryker Sustainability Solutions sent an "URGENT MEDICAL DEVICE RECALL" Customer Notification Letter to affected consignees asking them to follow the instructions in the letter. The letter also included a risk evaluation these devices may pose to end users and their patients. According to the Urgent Medical Device Recall Customer Notification Letter there are no harms associated with the use of affected reprocessed arthroscopic shavers. 1) Noncompatible devices are rendered inactive with console and alternative surgical methods are readily available. 2) Incorrect spin speed |
| Quantity | 14,265 units |
| Distribution | Nationwide Distribution: AL, AR, AZ, CA, CO, DC, DE, FL, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY and Puerto Rico. OUS Israel and Canada. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K012605 | REPROCESSED ATHROSCOPICS BURS Alliance Medical Corp., Nov 7, 2001 |