Recall Z-0976-2020 ยท FAD Stent, Ureteral
Cook Inc.
Sof-Flex Multi-Lenqth Ureteral Stent Set, Global Product No. G14773 G15000 G14867 G14951 G17852
| Date | Jan 4, 2019 |
| Status | Terminated |
| Reason | The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip. |
| Root cause | Process control |
| Action | On January 4, 2019, the firm distributed Urgent Medical Device Recall letters to customers via courier mail. Customers were informed that affected lots may contain a wire guide that was incorrectly loaded into the wire guide holder. |
| Quantity | 13388 total |
| Distribution | Domestic distribution to nationwide US. Foreign distribution to Australia, Belgium, Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India, Ireland, Italy, Japan, Korea, Luxembourg, Malaysia, |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.