ClearanceBrief
Recall Z-0964-2020 ยท OCY Endoscopic Guidewire, Gastroenterology-Urology

Cook Inc.

Wire Guide, Guidewire, Global Product Number G14326

DateJan 4, 2019
StatusTerminated
ReasonThe wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.
Root causeProcess control
ActionOn January 4, 2019, the firm distributed Urgent Medical Device Recall letters to customers via courier mail. Customers were informed that affected lots may contain a wire guide that was incorrectly loaded into the wire guide holder.
Quantity13388 total
DistributionDomestic distribution to nationwide US. Foreign distribution to Australia, Belgium, Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India, Ireland, Italy, Japan, Korea, Luxembourg, Malaysia,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082536COOK WIRE GUIDES
Cook Urological, Inc., Oct 27, 2008