ClearanceBrief
Recall Z-0996-2019 ยท MEC Disinfector, Medical Devices

Steris Corporation

AMSCO 5000 Series Washer/Disinfector Model # 5052

DateDec 20, 2018
StatusTerminated
ReasonThe software in the systems may not process the cycle originally intended. This could result in damage to the various medical devices and instruments processed in the washer/disinfector or potential for inadequate cleaning or disinfection.
Root causeSoftware Manufacturing/Software Deployment
ActionUrgent Device Field Correction notification letters dated 12/20/18 were sent to customers.
Quantity68
DistributionThe products were distributed to the following US states: AL, AR, CA, CO, FL, GA, IL, LA, MA, MD, ME, MI, NC, NJ, NM, NY, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, and WV. The products were distributed to the following foreign countries: Bahrain, Brazil, Canada, China, Saudi Arabia, South Korea, an
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.