Recall Z-0995-2019 ยท MEC Disinfector, Medical Devices

Steris Corporation

AMSCO 3000 Series Washer/Disinfector Model # 3052

DateDec 20, 2018
StatusTerminated
ReasonThe software in the systems may not process the cycle originally intended. This could result in damage to the various medical devices and instruments processed in the washer/disinfector or potential for inadequate cleaning or disinfection.
Root causeSoftware Manufacturing/Software Deployment
ActionUrgent Device Field Correction notification letters dated 12/20/18 were sent to customers.
Quantity11
DistributionThe products were distributed to the following US states: AL, AR, CA, CO, FL, GA, IL, LA, MA, MD, ME, MI, NC, NJ, NM, NY, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, and WV. The products were distributed to the following foreign countries: Bahrain, Brazil, Canada, China, Saudi Arabia, South Korea, an