Recall Z-1003-2019 ยท KQZ Component, Traction, Non-Invasive
Ossur Americas
Ossur, ReSolve Halo Vest, Part/ Description: 545700K/ ReSolve Halo Vest-shrt kodel; 545700L/ ReSolve Halo Vest-shrt lmbwool; 545800K/ ReSolve Halo Vest-tall kodel; 545800L/ ReSolve Halo Vest-tall lmbwool
| Date | Jun 21, 2018 |
| Status | Terminated |
| Reason | Halo system products labeled MR conditional will undergo confirmatory safety testing to confirm MR conditional parameters. In addition, the second line description on the titanium pin label mistakenly lists ceramic tip. The second line description should state titanium pin. |
| Root cause | Labeling design |
| Action | On 06/21/2018, Urgent Field Safety Notices were mailed to customers via UPS Next-Day Air requiring a signature of acceptance from the customer. Customers were advised the following: 1. Please pass this notice to those who need to be aware within your organization or to any department areas where the potentially affected devices will be used or transferred. 2. Maintain awareness on this notice for an appropriate period. 3. If you would like to have the product replaced please contact your customer service or sales representative and new product will be provided. Customers can exchange the affe |
| Quantity | 381 |
| Distribution | U.S.: HI, AL, CA, WA, AZ, TX, NM, NE, MI, IL, MN, MS, IN, OH, KY, TN, AK, AL, UT, SC, NC, GA, PA, FL, NY, DE, RI, VA, MA; Foreign: Australia, Germany, New Zealand, Belgium |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.