Recall Z-1256-2019 ยท DIC Alcohol Dehydrogenase, Specific Reagent For Ethanol Enzyme Method

Lin-Zhi International Inc

Ethyl Alcohol Enzymatic Assay, (a) REF 0220 (small test kit), (b) REF 0221 (large test kit) Product Usage: Detergent causes faster degradation in EtOH assay. Possible container or pitchers with some detergent residue was used while producing R2 of the defective lot.

DateJan 11, 2019
StatusTerminated
ReasonShelf life of the product may be reduced due to degradation of the assay.
Root causeMaterial/Component Contamination
ActionLin-Zhi International initiated the recall by email to its distributors on 01/11/2019. The letter identified the affected product, problem and actions to be taken. The distributors were directed to notify their customers and requested the destruction of the recalled product. For questions call T: ( 408) 970-8811 E: regulatory@lin-zhi.com
Quantity49 kits
DistributionUS Nationwide Distribution in the states of CA, MA, GA, KY, NC, MI, MD, NC, CT, AZ SC, TN, OR
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K032461ETHYL ALCOHOL ENZYMATIC ASSAY, CATALOG #0220 & 0221
Lin-Zhi International, Inc., Oct 3, 2003