Recall Z-1351-2019 ยท FSA Lift, Patient, Non-Ac-Powered

Moller Vital

Molift 2-Point sling bar, an accessory for Molift Mover 205. Item nos. 1430016 1430017 1430021

DateFeb 4, 2019
StatusOpen, Classified
ReasonThe firm has received 7 reports of malfunction for the device in which the hooks have broken off, potentially leading the sling to slip out of the sling bar and causing the patient to fall.
Root causeUse error
ActionOn February 4, 2019, the firm notified customers of the recall via Urgent Field Safety Notice letter. The letter alerted customers to the recalled product. The firm stated that all new 2-Point Sling Bars are now equipped with metal alloy hooks instead of the previously used composite hooks. As part of this recall, the firm will replace the recalled devices with the new design. Distributors were asked to do the following: 1) Read the instructions, and complete and return the customer response form 2) Notify affected customers of the recall 3) Review the list of sales records from Etac Supply
Quantity427 total
DistributionDomestic distribution to Arizona, Delaware, Florida, Georgia, Massachusetts, Ohio, South Carolina, Texas, Washington, and Wisconsin. Foreign distribution to Canada, Asia, Europe, and Australia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.