Recall Z-0963-2019 ยท LIT Catheter, Angioplasty, Peripheral, Transluminal

BrosMed Medical Co.,Ltd.

Hermes NC PTA Balloon Dilatation Catheter The balloon dilatation catheter is intended to dilate stenoses in the iliac, femoral, iliofemoral, popliteal, infrapopliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for stent dilatation post-deployment in the peripheral vasculature.

DateDec 14, 2018
StatusTerminated
ReasonIncorrect (higher) Rated Burst Pressure information printed on label.
Root causeEmployee error
ActionOn December 7, 2018 the firm, BrosMed Medical, sent an "URGENT FIELD SAFETY NOTICE" dated December 7, 2018 to their consignee. The notice described the product, problem and actions to be taken. The consignee was instructed to do the following: " Please reference the Attachment 1 for the list of Hermes NC PTA Balloon Dilatation Catheter with affected catalogue numbers and lots numbers. . Immediately stop selling the device from these lots. " Transfer this notification and inform relevant distributors/hospitals/clinics to check their inventory, complete and return the Effectiveness Check form
Quantity400
DistributionUS Distribution to MO.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K160941Castor, Achilles and Hermes NC PTA Balloon Dilatation Catheter
Brosmed Medical Co., Ltd., Dec 13, 2016