Recall Z-0969-2019 ยท MNL Accessories, Cleaning Brushes, For Endoscope

Madison Polymeric Engineering

Draco Enzymatic Deep-Cleaning Pads, Product Number HY0305Z Product Usage: Multi-tiered enzymatic detergent contains lipase, carbohydrase, amylase and protease. It effectively dissolves blood, fat, tissue, protein and most other forms of organic material Ready to use. Non-toxic, latex-free and neutral pH. Safe to use on all surgical instruments, flexible endoscopes and probes.

DateDec 14, 2018
StatusTerminated
ReasonThe product is contaminated with Pseudomonas.
Root causeNonconforming Material/Component
ActionCygnus sent an Urgent Medical Device Recall notification letters dated December 14, 2018 to customers. The letter identified the affected product, problem and actions to be taken. For questions contact Customer Service at 800-990-7489.
Quantity4800 pads
DistributionWorldwide Distribution - US states: CA, FL, IL, IN, NJ, NY, PA, and TN. The products were distributed to the following foreign countries: Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.