Recall Z-0769-2019 ยท LHI Set, I.V. Fluid Transfer

WEST PHARMA. SERVICES IL, LTD

Vial2BAG DC 20mm: 1) REF 6070112 & 2) REF 6070112C, Single use only, Rx Only

DateJan 28, 2019 (initiated Dec 26, 2018)
StatusTerminated
ReasonAdverse events have been reported with the use of oxytocin in conjunction with the 13mm Vial2Bag DC in pregnant women. Vial2Bag DC 20mm is recalled due to similarities with 13mm Vial2Bag DC.
Root causeDevice Design
ActionOn December 26, 2018, West Pharmaceutical Services, Inc. issued URGENT MEDICAL DEVICE CORRECTION and ACKNOWLEDGEMENT FORMS to their customers asking customers to avoid use of the drug oxytocin with the device. Actions to be taken by the Distributor: 1. Complete the required Acknowledgement and Receipt Form below and return (via email) to Kevin.Lentz@westpharma.com when complete (See attached form). 2. Prior to filling order / releasing shipments of any Vial2Bag DC 13mm products (which have been placed on hold per our request) to hospital customers, affix a copy of this Voluntary Field Corre
Quantity30 million units
DistributionUS Distributor in MO
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K170095Vial2Bag Direct Connect
Medimop Medical Project , Ltd., Mar 8, 2017