ClearanceBrief
Recall Z-1014-2019 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Siemens Healthcare Diagnostics, Inc.

Dimension Vista 500, Analyzer, chemistry Model no. 10284473

DateDec 12, 2018
StatusTerminated
ReasonWhen the Millipore Water Purification Module Progard Pretreatment Packs are used with Dimension Vista Triglyceride assay, patient and QC triglyceride results may be erroneously elevated as a result of contamination.
Root causeUnder Investigation by firm
ActionOn December 12, 2018, the firm issued Urgent Medical Device Correction letters to all Dimension Vista customers informing them of the issue. If customers have received a Progard Pretreatment Pack with a listed lot number, they are to follow the steps below upon Progard Pretreatment Pack installation and with use of the lots indicated. 1. Verify your water tank level is 85% or less on the WPM Maintenance screen (press Setup > Supplies > Maintain Water Supply > Maintenance). If your tank level exceeds 85%, use the Advanced->Diagnostics->Prime Pumps routine to lower the tank volume. 2. Replace
Quantity5373
DistributionNationwide domestic distribution. Worldwide foreign distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051087DIMENSION VISTA INTEGRATED SYSTEM
Dade Behring, Inc., Jul 11, 2005