Recall Z-0845-2019 ยท OSR Pacemaker/Icd/Crt Non-Implanted Components

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)

Medtronic Power Supply 26907, Medtronic CareLink Encore(TM) 29901 Programmer

DateJan 17, 2019
StatusTerminated
ReasonThere is a potential for an electrical short circuit to develop in a subset of 479 Model 26907 power supply cable plugs that connect to the Medtronic Model 29901 CareLink Encore programmer. A failure mode of the power supply has been found where an electrical short occurs on the low voltage direct current connector and results in the connector heating up, generating smoke, melting and in some cases disconnecting from the programmer.
Root causeDevice Design
ActionIn the US, beginning 17-Jan-2019, Medtronic Field Representatives (Field Reps) will locate all Encore programmers both in their trunk stock and in consignment locations. Since the Encore Power Supplies are interchangeable among all Encores, all Encore programmer distributed units are being located, and the associated Power Supplies will be checked to identify any affected units. If the Field Rep identifies an affected Power Supply physically at a customer account, the Medtronic Field Rep will hand deliver a courtesy FCA Notification Letter with instructions to the Risk Manager or HealthCare P
Quantity731 units
DistributionUS and Armenia, Bulgaria, Croatia, Georgia, Hungary, India, Iran, Jordan, Kazakhstan, Latvia, Lebanon, Oman, Pakistan, Qatar, Romania, Saudi Arabia, Syrian Arab Republic
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.