Recall Z-0795-2019 ยท OSR Pacemaker/Icd/Crt Non-Implanted Components

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)

MEDTRONIC CARELINK 2090 programmer

DateJan 17, 2019
StatusTerminated
ReasonThere is an error in how the programmer calculates and displays the remaining longevity value during a period of time (up to 2 years) prior to the device reaching its Recommended Replacement Time (RRT).
Root causeSoftware design
ActionIn the US, beginning 17-Jan-2019, Medtronic Field Representatives hand delivered a courtesy FCA Notification Letter to consignees who, according to Medtronic records, have programmers that are eligible for Micra software updates. Field Representatives will update the identified programmers using a provided USB. Notification will also be hand delivered to Micra Implanting or Follow-up physicians for awareness of the issue.
Quantity33007 units
DistributionUS, Armenia, Bulgaria, Croatia, Georgia, Hungary, India, Iran, Jordan, Kazakhstan, Latvia, Lebanon, Oman, Pakistan, Qatar, Romania, Saudi Arabia, Syrian Arab Republic
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.