ClearanceBrief
Recall Z-0807-2019 ยท NVZ Pulse Generator, Permanent, Implantable

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)

Medtronic implantable pulse generator: ATTESTA L DR MRI SureScan, Dual chamber rate responsive pacemaker (DDDR): (a) Model Number ATDR01 (b) Model Number ATDRL1 (c) Model Number ATDRS1

DateFeb 14, 2019 (initiated Jan 17, 2019)
StatusOpen, Classified
ReasonA subset of Medtronic dual chamber pacemakers distributed worldwide between 10 March 2017 and 7 January 2019 under the brand names Adapta, Versa, and Sensia when programmed to a dual chamber mode with atrial-sensing, may experience a circuit error that affects device functionality
Root causeComponent design/selection
ActionIn the US, beginning 17-Jan-2019, Medtronic Field Representatives hand deliver an FCA Notification Letter to consignees including implanting and follow-up physicians. In consultation with an Independent Physician Quality Panel (IPQP), patient management recommendations were provided with the FCA Notification letter to ensure patient safety. Medtronic Field Representatives will hand deliver an additional Supplemental Letter with the Urgent Medical Device Recall letter to a subset of physicians with patients whose device has shown evidence of a pacing pause that is potentially related to this c
Quantity9915 units
DistributionWordlwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.