ClearanceBrief
Recall Z-0828-2019 ยท ISY Joint, Knee, External Limb Component

Ottobock Orthopedic Industrie

21Y15 MagValve Product Usage: The PushValves/ MagValves is intended exclusively for the prosthetic fitting of lower limb amputees. The valve is a threadless one-way valve. It is used to evacuate air from the prosthetic socket.

DateJan 10, 2019
StatusTerminated
ReasonSome of the installed valve-inserts have a diameter smaller than the specification.
Root causeProcess control
ActionThe firm initiated the recall by email on 01/10/2019 and followed with a letter on 01/11/2019. The letter identified the affected product, problem, actions to be taken and that a second letter with the replacement unit was forthcoming.
Quantity110 units
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.