Recall Z-0766-2019 ยท NLH Catheter, Recording, Electrode, Reprocessed

Stryker Sustainability Solutions

BW Lasso 2515 ANV eco Variable Diagnostic EP Catheter, REF D134301 Diagnostic electrophysiology (EP) catheters are specially designed electrode catheters that transmit electrical impulses and can be positioned for endocardial recording or stimulation.

DateDec 17, 2018
StatusTerminated
ReasonStryker s Sustainability Solutions division (SSS) has received an increase in reports indicating that an EEPROM chip error code may occur when Reprocessed 2515 NAV eco Variable Electrophysiology (EP) Catheters are used with CARTO(R) EP Navigation Systems.
Root causeProcess change control
ActionThe firm, Stryker, sent an "URGENT MEDICAL DEVICE RECALL" Customer Notification Letter and attached Recall Effectiveness Check Form to Stryker Sustainability Solutions (SSS) sales representatives and international Stryker divisions to notify affected customers beginning on 12/17/2018.The letter described the product, problem and actions to be taken. The customers were instructed to do the following: Use of the affected product(s) should be discontinued. 1) Please check your inventory for the affected devices. Affected devices may be identified through Serial Number or Expiration Date. 2) Plea
Quantity103 units
DistributionUS distribution to states of: GA, KS, MO FL, CA, WA, IA,; and internationally to: Germany, Japan and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K112292REPROCESSED 2515 NAV VARIABLE ELECTROPHYSIOLOGY CATHETER
Stryker Sustainability Solutions, Oct 21, 2011