ClearanceBrief
Recall Z-0764-2019 ยท DYG Catheter, Flow Directed

Edwards Lifesciences, LLC

Edwards Lifesciences Swan Ganz Thermodilution Catheter, REF 131VF7P Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly determine hemodynamic pressures and cardiac output when used with a compatible cardiac output computer. Swan-Ganz thermodilution catheters are indicated for the assessment of a patient s hemodynamic condition through direct i

DateJan 31, 2019 (initiated Dec 21, 2018)
StatusTerminated
ReasonThese Swan Ganz catheters have incorrect lumen assembly, which causes reversal of the lumens. If the lumens are reversed the clinician may note reverse PA and CVP pressure values and waveforms.
Root causeNonconforming Material/Component
ActionAn Urgent Product Recall letter was sent via Fed-Ex to affected customers on 12/21/2018. A second communication was sent via Fed-Ex on 01/10/19. The letter identified the affected product, problem and actions to be taken. Customers were instructed to the return any unused units that are currently in their inventory. Once returned, replacement product will be shipped at no charge. Customers were asked to complete the attached acknowledgement form. For questions contact Technical Support at 1-800-822-8837 option 1.
Quantity9 units
DistributionWorldwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K160084Swan Ganz Catheters
Edwards Lifesciences, LLC, May 3, 2016