Recall Z-0755-2019 ยท HDA Forceps, Obstetrical

Fetzer Medical GmbH & Co. KG

16600 / Fetzer Surgical Luikart Simpson OB Forceps14" (35.5cm), Item No./Product Code 46-5050 Fetzer Medical Obstetrical Forceps is a hand-held instruments intended to assist the birth of a foetus during difficult vaginal births through grasping the foetal head in order to facilitate its passage through the birth canal. The instruments are only used for female population. The product is non-powd

DateJan 18, 2019 (initiated Dec 28, 2018)
StatusTerminated
ReasonThe firm had a report of doctors being unable to pull the obstetric forceps apart.
Root causeDevice Design
ActionThe recalling firm, Fetzer Medical, notified its consignees by email with an "URGENT: VOLUNTARY PRODUCT RECALL" letter dated 12/27/2018 on 12/28/20018. The letter described the product, problem and actions to be taken. The consignees were instructed to do the following: 1. Examine your inventory immediately to determine if you have product subject to this action on hand and quarantine such product(s). 2. Remove the product subject to this voluntary recall and communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be in
Quantity174 units
DistributionUS Distribution to states of: IL and VA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K172661Fetzer Medical Gynecological Forceps
Fetzer Medical GmbH & Co. KG, May 8, 2018