Recall Z-0802-2019 ยท IKZ Bed, Patient Rotation, Powered

Arjohuntleigh Magog

ARJOHUNTLEIGH GETINGE GROUP, RotoProne Therapy System, Model Number 209800

DateFeb 5, 2019 (initiated Dec 19, 2018)
StatusTerminated
ReasonInability or difficulty in unbuckling/releasing the proning arm buckle might delay patient access in an emergency situation such as cardiac arrest. This, in turn, could contribute to a life-threatening event.
Root causeComponent design/selection
ActionArjo hand-delivered the notice with each RotoProne Therapy System rental and delivery beginning on 12/19/2018. Arjo will require the customer to confirm the notification has been received and understood via the signed delivery notification.
Quantity231 units
DistributionUS, Saudi Arabia, and Kuwait
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.