Recall Z-0745-2019 ยท DXT Injector And Syringe, Angiographic

Bayer Medical Care, Inc.

MAVIG Portegra or Portegra2 Overhead Counterpoise System (OCS) used in conjunction with the Medrad Stellant CT Injection System (Certegra Workstation), Medrad Mark 7 Arterion" Injection System, or Medrad Mark V ProVis Injection System, Catalog Numbers: Catalog numbers: OCS CEIL (58-S, 58-P, 85-S, 85-P, 100-S, 100-P, 70Si, 97Si), OCS TRACK (58-S, 58-P, 80-S, 80-P), OCS WALL (S, P), OCS (115A, 115C

DateDec 14, 2018
StatusTerminated
ReasonIf the head mount knob is not properly tightened and the operator directly uses the injector head to move the unit, it may become disengaged, causing the injector head to become loose or fall.
Root causeDevice Design
ActionUrgent Medical Device Field Corrective Action notification letters were sent to customers on 12/12/18.
Quantity16656
DistributionThe products were distributed US nationwide. The products were distributed to the following foreign countries: Algeria, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Belgium, Bermuda, Brazil, Brunei Darussalam, Canada, Chile, China, Cote D'Ivoire, Cuba, Curacao, Cyprus, Czech Republic, Denmark
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K023183MEDRAD STELLANT CT INJECTOR SYSTEM
Medrad, Inc., Jan 10, 2003
K112086MEDRAD MARK 7 ARTERION (PEDESTAL SYSTEM / TABLE MOUNT SYSTEM), MEDRAD MARK 7 ARTERION SYRINGE
Medrad, Inc., Oct 14, 2011
K173773MEDRAD Stellant FLEX CT Injection System with Certegra Workstation, MEDRAD Stellant FLEX Syringe Kits, MEDRAD Stellant C
Bayer Medical Care, Inc., Aug 17, 2018
K903390MEDRAD MARK V-PLUS INJECTOR
Medrad, Inc., Sep 26, 1990