Recall Z-1124-2019 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Roche Diagnostics Corporation

COBAS INTEGRA 400 plus analyzer used in the following assays: a)Tina-quant Apolipoprotein A-1 ver.2 Catalog Number: 03032566122 b)Tina-quant Albumin Gen.2 Catalog Number 04469658190 c)Creatinine plus ver.2 Catalog Number 03263991190 d) riglycerides Catalog Number :20767107322 - Product Usage: Tina-quant Apolipoprotein A-1 ver.2 A lipoprotein test system is a device intended to measure lipo

DateDec 14, 2018
StatusTerminated
ReasonCOBAS INTEGRA 400 plus and cobas c 111 analyzers Additional Extra Wash Cycle Configurations and/or Batch Mode Needed for Specific Tests Due to Carry-over
Root causeDevice Design
ActionRoche Diagnostics issued on 12/14/18 "Urgent Medical Device Correction (UMDC) via UPS Ground (receipt required) Consignee Instructions: Manually add the EWCs listed in the Urgent Medical Device Correction (UMDC) on your COBAS INTEGRA 400 plus and cobas c 111 analyzers. If customers run the test combinations requiring needle assignment (See Table on page 3 of UMDC TP-00508), run those in batch mode until the updated TAS has been loaded on the COBAS INTEGRA 400 Plus analyzer. Contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you
Quantity656 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K013249TINA-QUANT APOLIPOPROTEIN VER.2
Roche Diagnostics Corp., Nov 16, 2001
K101203TINA-QUANT ALBUMIN GEN 2
Roche Diagnostics, Sep 10, 2010