Recall Z-0733-2019 ยท LZO Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented

Waldemar Link GmbH & Co. KG (Mfg Site)

LINK STEM with Microporous Surface MP Reconstruction Prosthesis, CEMENTLESS (Size XS-0, D: 12mm, L: 160mm), Reference Number 172-916/12 Product Usage: The LINK STEM with Microporous Surface MP Reconstruction Prosthesis, CEMENTLESS System is indicated for patients with mobility-limiting diseases, fractures or defects which cannot be treated by conservative or osteosynthetic procedures. The LINK ST

DateDec 7, 2018
StatusTerminated
ReasonThe reaming guide and the guide rod can't be screwed into the intended depth. The reason for this is a reduced depth of the core hole in the stem. A reaming guide that is not screwed in completely may lead to a proximal preparation in the depth that is not sufficient and to a neck that can't be positioned correctly. The correct fit of the neck can't be verified by means of the guide rod.
Root causeUnder Investigation by firm
ActionUrgent Medical Device Recall notification letters dated 12/7/18 were distributed to customers. Please identify your total inventory for this product. Please take the following actions: 1) Immediately cease use and segregate the above described product. 2) Complete the enclosed Consignee Reply Form, even if you have no inventory. 3) Return the above product and a copy of the Consignee Reply Form using Federal Express Account #: 007183879 to: LinkBio Corporation 101 Roundhill Drive Rockaway, NJ 07446 ATT: Nikhil Mangale Recall Return Authorization #: R-2018-03
Quantity11
DistributionUS Nationwide in the states of: FL, GA, IA, IN, KS, MI, MN, and TN.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K142187LINK MP RECONSTRUCTION PROSTHESIS