Recall Z-0696-2019 ยท GDW Staple, Implantable

Ethicon Endo-Surgery Inc

PROXIMATE PPH Hemorrhoidal Circular Stapler Product Code:PPH03 PPH03: The PROXIMATE PPH Hemorrhoidal Circular Stapler and Accessories have application throughout the anal canal to perform surgical treatment of hemorrhoidal disease

DateDec 7, 2018
StatusTerminated
ReasonHemorrhoidal Circular Stapler may have been assembled without a washer, this may potentially cause bleeding or soft tissue injury during use as the device may not fully cut when fired
Root causeProcess control
ActionThe firm, Ethicon, issued an "URGENT: MEDICAL DEVICE RECALL (REMOVAL)" to US consignees on 12/7/18, via UPS 2nd day mail. Ethicon Johnson & Johnson Affiliates in North America (Canada), Europe, LATAM, and Asia Pacific will notify their impacted consignees by providing the customer letter through post, email or sales representative visits. Letter identifies reason for recall, action to be taken: examine inventory and return recalled product for replacement; complete the Business Reply Form (BRF) confirming receipt of this notice and fax or email it to Stericycle at 1-844-524-3888 or Ethicon8
Quantity15,264 units
DistributionWorldwide Distribution: US (Nationwide) and to countries of: ARGENTINA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, ECUADOR, EGYPT, INDIA, INDONESIA, JAPAN, KOREA, KUWAIT, LEBANON MEXICO, NICARAGUA, PAKISTAN, SAUDI ARABIA, SINGAPORE, TAIWAN URUGUAY and VIETNAM.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051301PROXIMATE PPH HEMORRHOIDAL CIRCULAR STAPLER AND ACCESSORIES
Ethicon Endo-Surgery, Inc., Aug 30, 2005