ClearanceBrief
Recall Z-0700-2019 ยท OAD Catheter, Percutaneous, Cardiac Ablation, For Treatment Of Atrial Flutter

Boston Scientific Corporation

IntellaTip MiFi Open-Irrigated Ablation Catheter Product Usage: Open Irrigated ablation catheters, when used with a compatible Radiofrequency (RF) Controller and Irrigation Pump, are indicated for: Cardiac electrophysiological mapping Delivering diagnostic pacing stimuli RF ablation of sustained or recurrent type 1 Atrial Flutter in patients age 18 years or older Treatment of drug refract

DateNov 28, 2018
StatusOpen, Classified
ReasonSupplier had changed its resin sub-suppliers three (3) times without notifying BSC. These changes resulted in slight variations in resin material compositions which were not reflected in BSC s internal product documentation. Not approved in customer document.
Root causeNonconforming Material/Component
ActionOn November 28, 2018, Boston Scientific sent Urgent Medical Device Removal - Immediate Action Required letters to their Japanese consignees requesting them to check and segregate any of the recalled product. The letter also asked the consignees to conduct sub-recalls, if necessary, and to return The Reply Verification Tracking Form.
Quantity1,282 (all three types in Japan)
DistributionJapan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.