Recall Z-0699-2019 ยท OAD Catheter, Percutaneous, Cardiac Ablation, For Treatment Of Atrial Flutter

Boston Scientific Corporation

Blazer Open-Irrigated Ablation Batch Catheter Product Usage: Open Irrigated ablation catheters, when used with a compatible Radiofrequency (RF) Controller and Irrigation Pump, are indicated for: Cardiac electrophysiological mapping Delivering diagnostic pacing stimuli RF ablation of sustained or recurrent type 1 Atrial Flutter in patients age 18 years or older Treatment of drug refractor

DateNov 28, 2018
StatusOpen, Classified
ReasonSupplier had changed its resin sub-suppliers three (3) times without notifying BSC. These changes resulted in slight variations in resin material compositions which were not reflected in BSC s internal product documentation. Not approved in customer document.
Root causeNonconforming Material/Component
ActionOn November 28, 2018, Boston Scientific sent Urgent Medical Device Removal - Immediate Action Required letters to their Japanese consignees requesting them to check and segregate any of the recalled product. The letter also asked the consignees to conduct sub-recalls, if necessary, and to return The Reply Verification Tracking Form.
Quantity1,282 (all three types in Japan)
DistributionJapan
FDA recordOpen on fda.gov