ClearanceBrief
Recall Z-0850-2019 ยท FKX System, Peritoneal, Automatic Delivery

Baxter Healthcare Corporation

Kaguya Automated Peritoneal Dialysis System Product Usage: It is intended for automatic control of dialysate solution exchanges in the treatment of adult renal failure patients undergoing Peritoneal dialysis.

DateDec 11, 2018
StatusTerminated
ReasonPotential for the software on Automated PD System cyclers which can cause shortened dwell times during a Cycle-Based therapy
Root causeSoftware design
ActionBaxter Healthcare Corporation notified providers on about 12/11/2018 of the recall concerning the AMIA Automated Peritoneal Dialysis System, via "Urgent Medical Device Correction" letter. Patients were notified on about 1/14/2018 via "Urgent Medical Device Correction" letter. Providers were instructed to continue using the AMIA Automated PD System however, for patients who are prescribed a Cycle-Based therapy, to change these therapies to Time-Based by following the programming instructions within the Clinician Guide, Section 5.3.1 - Time-Based Programming Method. In Time-Based therapy, t
Quantity1,085 devices
DistributionWorldwide Distribution - US Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. Distributed internationally to Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K151525Amia Automated PD System with Sharesource
Deka Research and Development, Oct 9, 2015