ClearanceBrief
Recall Z-0704-2019 ยท KPS System, Tomography, Computed, Emission

GE Healthcare, LLC

VariCam

DateOct 30, 2017
StatusCompleted
ReasonA potential hazardous situation may occur if the collimator locking handle is not properly placed in the lock position and the warning circuit does not detect that the collimator is unlocked because of the patient proximity to the collimator during a clinical scan.
Root causeOther
ActionThe firm initiated the medical device correction by letter on 10/0/2017. The letter instructed the consignee to ensure the collimator locking handles are in the locked position at all times. GE Healthcare will inspect and if necessary correct all affected products.
Quantity64 units
DistributionWorldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K953801VARICAM
Elscint, Inc., Nov 22, 1995