Recall Z-0785-2019 ยท CAC Apparatus, Autotransfusion

Medtronic Perfusion Systems

RESERVOIR "Y" ADAPTER, (a) BT920 (b) BT926 Suction and Anticoagulation assemblies are indicated for use with an autotransfusion system between the suction tip and reservoir.

DateNov 27, 2018
StatusOpen, Classified
ReasonMedtronic discovered that certain Y-connectors included in the suction lines have exhibited tearing and cracking. These defects could cause the Y-connectors to leak when in use, which could result in potential blood loss and a breach in the circuit.
Root causeProcess control
ActionThe firm, Medtronic, began notifying US consignees on 27-Nov-2018 via hand delivery to be followed up with a UPS 2-Day communication titled "URGENT MEDICAL DEVICE RECALL" to each listed consignee that does not confirm hand delivery. Communications OUS also began 27-Nov-2018 using locally approved methods. The letter described the product, problem and actions to be taken. Customers were requested to do the following: 1. Identify and quarantine all unused affected product as listed in your inventory. 2. Return all unused affected product in your inventory to Medtronic. Contact Medtronic Cust
Quantity1100 units
DistributionWorldwide Distribution: US (nationwide) and countries of: Australia, Canada, France, Germany, Japan, and Sweden.