Recall Z-0709-2019 ยท LXT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component, Metal Composite

Zimmer Biomet, Inc.

Biomet DVR Crosslock Screws- 2.7mm x18mm, Sterile ITEM 131827118 Product Usage: Usage: The system is intended for stabilization and fixation of small bone fragments in fresh fractures, revision procedures, joint fusion and reconstructions (osteotomies) of small bones of the hand, foot, radius, ulna, ankle, humerus, scapula and pelvis, particularly in osteopenic bone

DateNov 20, 2018
StatusTerminated
ReasonOne lot of product labeled 2.7mm x18mm Lock Screw Square Drive (Lot 029970) contains 20mm locking screws and one lot of product labeled 2.7mm x 20mm Lock Screw Square Drive (Lot 029980) contains 18mm locking screws
Root causePackaging process control
ActionZimmer Biomet issued Urgent Medical Device Recall letter on 11/20/18 advising of the problem, heath risk and action to be taken- Quarantine and return product. Complete Attachment 1 Certificate of Acknowledgement. Contact customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday.
Quantity50 pieces
DistributionCountry of Japan