Recall Z-0708-2019 ยท LXT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component, Metal Composite
Zimmer Biomet, Inc.
Biomet DVR Crosslock Screws- 2.7mm x 20mm, Sterile ITEM 131827120 Product Usage: The system is intended for stabilization and fixation of small bone fragments in fresh fractures, revision procedures, joint fusion and reconstructions (osteotomies) of small bones of the hand, foot, radius, ulna, ankle, humerus, scapula and pelvis, particularly in osteopenic bone
| Date | Nov 20, 2018 |
| Status | Terminated |
| Reason | One lot of product labeled 2.7mm x18mm Lock Screw Square Drive (Lot 029970) contains 20mm locking screws and one lot of product labeled 2.7mm x 20mm Lock Screw Square Drive (Lot 029980) contains 18mm locking screws |
| Root cause | Packaging process control |
| Action | Zimmer Biomet issued Urgent Medical Device Recall letter on 11/20/18 advising of the problem, heath risk and action to be taken- Quarantine and return product. Complete Attachment 1 Certificate of Acknowledgement. Contact customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. |
| Quantity | 50 pieces |
| Distribution | Country of Japan |
| FDA record | Open on fda.gov |