Recall Z-0685-2019 ยท MUE Full Field Digital, System, X-Ray, Mammographic

Hologic, Inc.

I-View Contrast Enhanced Digital Mammography, ASY-08109

DateDec 6, 2018
StatusTerminated
ReasonCalibration issue not possible to visualize contrast uptake in the subtracted mammography images
Root causeSoftware design
ActionHologic issued letter on 12/6/2018 USPS Certified Mail to customers to inform them of this product correction, and Hologic asks customers to only use I-View for patients with known lesions and confirm the image quality is of acceptable diagnostic quality.Hologic will scheduling appointments to recalibrate the systems for contrast enhanced digital mammography.
Quantity337 units
DistributionNationwide Foreign: Australia Austria Belgium Brazil Canada China Colombia Ecuador Egypt France Germany Greece Hong Kong India Ireland Israel Italy Japan Korea, Republic of Kuwait Malaysia Mexico Netherlands Pakistan Peru Poland Portugal Saudi Arabia Singapore Slovakia South Africa Spain Switzerland
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K123873CONTRAST ENHANCED DIGITAL MAMMOGRAPHY
Hologic, Inc., Jan 29, 2013