Recall Z-0970-2019 ยท GEH Unit, Cryosurgical, Accessories
PENTAX of America Inc
C2 CryoBalloon Controller, REF: FG-1017, with Controller Software v1.18.258 The C2 CryoBalloon Ablation System is intended to be used as a cryosurgical tool in the field of general surgery, specifically for endoscopic application and the ablation of dysplastic Barrett s Esophagus.
| Date | Dec 3, 2018 |
| Status | Terminated |
| Reason | The Controller does not detect overpressure in the balloon during the application of non-dosing puffs of Nitrous Oxide, which can contribute to balloon over pressurization, if the intended vent lumen of the catheter is significantly occluded to prevent relieving balloon pressure due to a kinked catheter condition.If a patient is exposed to higher than physiologic pressures, adverse events such as perforation or mucosal laceration may occur. |
| Root cause | Nonconforming Material/Component |
| Action | On 10/18/18, the firm, Pentax Medical, started contacting US Affected Customers by telephone to inform them of the pending action and to advise they discontinue use and quarantine affected devices. On 12/03/18, "URGENT MEDICAL DEVICE REMOVAL" letters dated 11/16/18 were mailed via USPS Certified Mail to its customers. Customer response forms and Customer Return Material Shipment Labels were mailed with the removal notices. Customers were informed that affected products should not be used and should be returned to the firm. In addition, customers were asked to ensure that all potential users in |
| Quantity | 45 |
| Distribution | Worldwide Distribution: US (nationwide) to states of:: CA, DC, FL, LA, MA, MD, MN, NC, NJ, NY, OH, PA, SC, TN, TX, VA, WA;; and countries: Canada, Germany, and Netherlands. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K163684 | C2 CryoBalloon Ablation System C2 Therapeutics, Inc., Jan 23, 2018 |