ClearanceBrief
Recall Z-0679-2019 ยท MUJ System, Planning, Radiation Therapy Treatment

RAYSEARCH LABORATORIES AB

RayStation 3.5, RayStation 4.0, RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6, RayStation 7 and RayStation 8A

DateNov 12, 2018
StatusTerminated
ReasonIsocenter shifts when using dose tracking in RayStation 3.5, RayStation 4.0, RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6, RayStation 7 and RayStation 8A. In these versions, isocenter shifts in dose tracking are incorrectly interpreted according to the DICOM patient coordinate system.
Root causeUnder Investigation by firm
ActionRayStation issued Field Safety Notice, Medical Device Correction #37483, on 11/12/18 via email, states issue, action to be taken and correction of software in Dec 2018.
Quantity69
DistributionAZ CA CO CT DE FL GA HI ID IL LA MA MD ME MI MO MS NC NJ NY OH OR PA RI TN TX VA VT WA WI
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K180379RayStation 7
RaySearch Laboratories AB (PUBL), Apr 25, 2018