ClearanceBrief
Recall Z-2138-2019 ยท PGQ Cochlear Implant With Combined Electrical Stimulation And Acoustic Amplification

MED-EL Elektromedizinische Gereate, Gmbh

MED EL Cochlear Implant System, PULSARci100 Standard - Product Usage: The PULSARci100 Implant is an implantable part of the MED EL Cochlear Implant System and can only be used together with compatible MED EL external components. The MED EL Cochlear Implant System is intended to evoke auditory sensations via electrical stimulation of the auditory pathways for severely to profoundly hearing impaired

DateOct 16, 2018
StatusTerminated
ReasonDevices were distributed despite the Helium-fine leak test results not being within the cochlear implant standards specification.
Root causeEmployee error
ActionMed-EL initially notified customers who had not yet implanted the affected devices on about 10/16/2018 via telephone call and followed up with an "Urgent-Medical Device Field Safety Notice (FSN)." The notice instructed customers to not implant the device if still available in inventory. Customers who had implanted the device were notified on about 11/27/2018 via telephone call and followed up with an "Urgent-Medical Device Field Safety Notice (FSN)." The notice instructed customers to immediately inform affected clinics of the recall and the long term risk. MED-EL does not recommend a de
Quantity1 device
DistributionWorldwide distribution - US Nationwide in the states of California and Illinois. Countries of ARGENTINA, AUSTRALIA, AUSTRIA, BRAZIL, BULGARIA, CHINA, FRANCE, GEORGIA, KOREA, REPUBLIC OF, POLAND, RUSSIAN FEDERATION, SPAIN, SWEDEN.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.