Recall Z-0687-2019 ยท LOM Test, Hepatitis B (B Antigen, B Surface Antigen, Be Antigen)

ORTHO-CLINICAL DIAGNOSTICS

VITROS Anti-HBe Reagent Pack, Cat No. 6801816 Product Usage: For the in vitro qualitative detection of antibodies to hepatitis B e antigen (anti-HBe) in human adult and pediatric (2 to 21 years old) serum from individuals who have symptoms of chronic hepatitis and those who have recovered from HBV infection, using the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic Syst

DateNov 1, 2018
StatusOpen, Classified
ReasonThe firm is extending their previous July 2018 recall to include additional product codes. Biased results may occur for these additional products at biotin concentrations which are lower than indicated in the current Instructions for Use.
Root causeDevice Design
ActionOn November 1, 2018, a customer letter was sent via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail to all customers who could process the affected products to inform them of the potential interference of biotin at levels lower than stated in the IFUs and that Ortho is in the process of revising the IFUs for the affected products. Ortho recommended customers follow normal laboratory procedures for troubleshooting of samples containing other assay interferences and it was acceptable to continue using the affected products. If you have question
Quantity3429 units (100 tests per unit)
DistributionWorldwide Distribution - US Nationwide domestic distribution. International distribution to Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, India, France, Japan, Germany, Italy, Mexico, Norway, Poland, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, and Uni
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.