ClearanceBrief
Recall Z-0608-2019 ยท KPS System, Tomography, Computed, Emission

GE Healthcare, LLC

Discovery NM/CT 670 Pro, model 5376204-70-54

DateMay 24, 2018
StatusCompleted
ReasonWhen your system was installed, the Main disconnect switch, EPO button on the A1 panel, and Remote EPO button(s) (outside of the A1 panel) may not have been wired correctly which could result in these not shutting off all power to the entire system as intended, resulting in a potential electric hazard.
Root causeComponent change control
ActionThe firm initiated the recall by letter on 05/24/2018. The letter explained the issue and requested the UPS(s) be disconnected from the system until the system could be repaired.
Quantity1 unit
DistributionCA, DC, OH, UT, WI, Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K093514DISCOVERY MODEL NM/CT 670
Ge Medical Systems F.I. Haifa, Dec 10, 2009