ClearanceBrief
Recall Z-0634-2019 ยท KGO Surgeon'S Gloves

Cardinal Health 200, LLC

Protexis Neoprene Surgical Glove, Size 7.0 This powder-free sterile light brown colored surgeons gloves are a disposable device made of synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.

DateOct 31, 2018
StatusTerminated
ReasonLabeling issue in that the breakthrough (i.e., permeation) times for two chemotherapy drugs (Carmustine and Thiotepa) stated on the wallet (primary packaging) of the Protexis Neoprene surgical gloves are incorrect. Additionally, breakthrough times for Mitomycin were added to the same label in error.
Root causeLabeling mix-ups
ActionCardinal Health notified customers on about 10/31/2018 via "URGENT: PRODUCT CORRECTION NOTICE." Customers were instructed to examine inventory for affected products, affix the provided warning label on the principal display panel on the affected dispenser packs, notify internal parties and/or customers if the product was further distributed of the recall, and complete and return the provided Acknowledgement Form. If the wallets containing the gloves are removed from the dispenser packs, customers were also instructed to provide a warning label to accompany the affected product inform users o
Quantity1,475,095 pairs total
DistributionWorldwide Distribution: US (Nationwide distribution) to states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and to countries of: Austria, Belg
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K113707STERILE NEOPRENE POWDER-FREE SURGICAL GLOVES WITH NITRILE COATING
Cardinal Health-Medical Products and Services, Apr 27, 2012