Recall Z-0657-2019 ยท LXH Orthopedic Manual Surgical Instrument
Howmedica Osteonics Corp.
Reliance LITe Nerve Hook Bayoneted, 90, Blunt Tip, REF # 48066103 The blunt probe can be used as a conventional means of achieving nerve retraction. The bend at the proximal tip of the ring the discectomy, as well as during the insertion of an IBD during a conventional TLIF procedure.
| Date | Nov 26, 2018 |
| Status | Terminated |
| Reason | Instead of having a spherical diameter, the probe was observed to have 2 radii and a flat surface. Additionally, upon further investigation, it was observed that some parts exhibited an edge on the intersection between the radius and the diameter of the probe. |
| Root cause | Under Investigation by firm |
| Action | The firm, Stryker, sent an "Urgent Medical Device Recall" letter dated 11/14/2018 to its customers on 11/26/2018. The letter described product, problem and actions to be taken. The customers were instructed to: Examine your inventory; Immediately quarantine the affected product; and complete customer response form to Stryker Spine at email: Spine-RegulatoryAction@stryker.com. For questions contact Manager-Regulatory Compliance & Quality System at 866-45-SPINE (77463).. |
| Quantity | 39 |
| Distribution | Worldwide distribution: US Distribution to the state of: PA and countries of:: Australia, Canada, France, Italy, Netherlands, Poland, and Spain. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.