ClearanceBrief
Recall Z-0644-2019 ยท LZO Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented

Zimmer GmbH

Avenir Muller Stem Standard, uncemented HA, 1, Taper 12/14 Item Number: 01.06010.001 - Product Usage: Avenir@ Muller Stems are intended to reduce pain and increase hip mobility through long term cementless fixation of total or hemi hip arthroplasty in the femur of patients with an adequate bone stock to support the component. A system consisting of a stem, a ball head and a cup is used for the tre

DateOct 31, 2018
StatusTerminated
ReasonMislabeled Labeled Avenir Mller Stem size 1, the stem inside the packaging is an Avenir Mller Stem size 2 .
Root causeProcess control
ActionZimmer Biomet issued Notices to the OUS consignees on October 31, 2018. Letter identifies action to be taken: Review this notice and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. Complete Attachment 1 Certificate of Acknowledgement
Quantity13 units
DistributionInternational distribution in the countries of Germany and Switzerland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K123392AVENIR MULLER STEM
Zimmer GmbH, Mar 4, 2013