Recall Z-0551-2019 ยท IZD Probe, Uptake, Nuclear

Capintec Inc

Captus 700 Thyroid Uptake System, Model # 5430-3137 and 5430-3138

DateDec 4, 2018 (initiated Nov 5, 2018)
StatusTerminated
ReasonThere is a potential for the spring arm failure as a result of a broken tension rod (a component inside the arm). A break in the tension rod can casue the collimator to fall downward to its lowest point of travel, approximately 25 inches from the ground.
Root causeUnder Investigation by firm
ActionNotification letters were sent to customers on 11/5/18. A Revised Product Alert - Corrective Action letter was sent to customers beginning 1/15/19 due to an expansion of the event.
Quantity65
DistributionWorldwide distribution - US nationwide distributed in the states of AR, AZ, CA, CO, CT, DC, FL, IA, IN, LA, MD, MI, MN, NC, NJ, OH, OK, PA, SC, SD, TX, VA, WI, WV, and countries of Argentina, Brazil, Canada, China, Cyprus, Greece, Hong Kong, India, Japan, South Korea, Vietnam.