ClearanceBrief
Recall Z-0550-2019 ยท IZD Probe, Uptake, Nuclear

Capintec Inc

Captus 3000 Thyroid Uptake System, Model # 5430-0076 and 5430-0077

DateDec 4, 2018 (initiated Nov 5, 2018)
StatusTerminated
ReasonThere is a potential for the spring arm failure as a result of a broken tension rod (a component inside the arm). A break in the tension rod can casue the collimator to fall downward to its lowest point of travel, approximately 25 inches from the ground.
Root causeUnder Investigation by firm
ActionNotification letters were sent to customers on 11/5/18. A Revised Product Alert - Corrective Action letter was sent to customers beginning 1/15/19 due to an expansion of the event.
Quantity385
DistributionWorldwide distribution - US nationwide distributed in the states of AR, AZ, CA, CO, CT, DC, FL, IA, IN, LA, MD, MI, MN, NC, NJ, OH, OK, PA, SC, SD, TX, VA, WI, WV, and countries of Argentina, Brazil, Canada, China, Cyprus, Greece, Hong Kong, India, Japan, South Korea, Vietnam.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.