Recall Z-0676-2019 ยท CKG Radioimmunoassay, Acth

Tosoh Bioscience Inc

ST AIA-PACK ACTH (ACTH: Adrenocorticotropic Hormone) is designed for IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of Adrenocorticotropic hormone (ACTH) in human EDT A plasma. Measurements of ACTH are used in the differential diagnosis and treatment of certain disorders of the adrenal glands such as Cushing's syndrome, adrenocortical insufficiency and the ectopic ACTH syndrome.

DateNov 30, 2018
StatusTerminated
Reason
Root causeDevice Design
ActionOn November 30, 2018, Tosoh Bioscience issued Urgent Medical Device Recall notices to customers. Customers are advised to take the following actions: " Do not use the assays identified in notice for those patients who have recently had a fluorescein angiography procedure. For these patients, use an alternative testing method that does not use a fluorescein-based technology. " Inform medical professionals that the Tosoh AIA assays in the notice utilize a fluorescein-based technology and should not be used to test patient samples for patients who have recently undergone fluorescein angiograph
Quantity1085 units
DistributionUS Nationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K111335ST AIA-PACK ACTH, AND ST AIA-PACK ACTH CALIBRATOR SET MODEL 025221 AND 025321
Tosoh Bioscience, Inc., Dec 1, 2011