ClearanceBrief
Recall Z-0675-2019 ยท LPS Urinary Homocystine (Nonquantitative) Test System

Tosoh Bioscience Inc

ST AIA-PACK Homocysteine (HCY) is designed for IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of homocysteine in human serum, heparinized plasma or EDT A plasma on TOSOH AIA System Analyzers. Homocysteine measurements are used in the diagnosis and treatment of hyperhomocysteinemia or homocysteinuria.

DateNov 30, 2018
StatusTerminated
Reason
Root causeDevice Design
ActionOn November 30, 2018, Tosoh Bioscience issued Urgent Medical Device Recall notices to customers. Customers are advised to take the following actions: " Do not use the assays identified in notice for those patients who have recently had a fluorescein angiography procedure. For these patients, use an alternative testing method that does not use a fluorescein-based technology. " Inform medical professionals that the Tosoh AIA assays in the notice utilize a fluorescein-based technology and should not be used to test patient samples for patients who have recently undergone fluorescein angiograph
Quantity992 units
DistributionUS Nationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K121533ST AIA-PACK HOMOCYSTEINE; ST AIA-PACK HOMOCYSTEINE CALIBRATOR SET; AIA-PACK HOMOCYSTEINE CONTROL SET
Tosoh Bioscience, Inc., Jun 15, 2012